BRAND Transferpette S-12 12 Channel Adjustable Volume Automatic Pipette
Description;
BRAND's Transferpette® S is the same high-quality pipette customers have come to expect. A lightweight yet robust instrument offering a high degree of accuracy and precision, the Transferpette® S now offers newly enhanced features that make working easier and more efficient.
- With its new handle geometry, Transferpette® S provides a comfortable grip regardless of hand size and when working with the right or left hand. Also, true one-handed operation means that the grip does not change or the sample is placed on the ground to adjust the volume level.
- The newly designed volume setting brake is easy to use and clearly shows that the pipette is secured against accidental volume changes.
- An even thinner shaft makes working with narrow tubes and bottles more manageable - save more of your sample and the shaft coupling to the handle has been redesigned to improve ergonomics and make cleaning easier.
- Color-coded pipette highlights for tip selection are even clearer.
Transferpette® S Single Channel Fixed Volume and Transferpette® S 8- and 12- Multichannel Pipettes are also available.
- One-handed volume adjustment with brake to protect against accidental volume changes
- 12.5 mm piston stroke length is the shortest of any leading pipette
- Easy Calibration™ technology provides fast, easy, in-lab calibration adjustments without tools
- The four-digit volume display is always visible
- Universal tip cone fits most premium manufacturers' tips
- Fully and repeatedly autoclavable at 121 °C / 250 °F (20 minutes)
Transferpette® S Single Channel Pipettes are supplied with performance certificate, instruction manual, rack mount and a two-year warranty.
Usage Limitations - This device is designed for pipetting liquids within the following limitations:
- Use instrument and reagent between +15 °C and +40 °C (59 °F to 104 °F)
- Steam pressure up to 500 mbar
- Viscosity: 260 mPa/s (260 cps)
Exception Handling- This instrument cannot be used with:
- Liquids incompatible with polypropylene
- Liquids not compatible with polycarbonate (view window)
- Liquids incompatible with FKM and polyether ether ketone (PEEK)
- Liquids incompatible with polyvinylidene fluoride
- Liquids incompatible with polyphenylsulfite (PPS) (in an adjustable 50 µl instrument)
- Liquids with very high vapor pressure
Technical data;
Code number | Capacity | Model | A* ≤ ± % | CV* ≤ % | Lower Limit |
705920 | 0,5-10 µL | M12-10 | 1.6 | 1.0 | 0,01 µL |
705926 | 5-50 µL | M12-50 | 0.8 | 0.4 | 0.05 µL |
705928 | 10-100 µL | M12-100 | 0.8 | 0,3 | 0.1 µL |
705930 | 20-200 µL | M12-200 | 0.8 | 0.3 | 0.2 µL |
705932 | 30-300 µL | M12-300 | 0.6 | 0.3 | 0.5 µL |
A*=Accuracy, CV*=Coefficient of Variation
*Accuracy value and coefficient of variation are final test values referring to distilled water with given nominal volume, device and ambient temperature (20°C/68°F) in equilibrium and smooth operation.